A technique for pediatric total skin electron irradiation
Qinan Bao1, Brian A Hrycushko, Joseph P Dugas
1Department of Radiation Oncology, University of Texas Southwestern Medical Center, 5801 Forest Park Rd, Dallas, TX 75390, USA. qinan.bao@utsouthwestern.edu
Total skin electron irradiation (TSEI) was adapted for pediatric patients with leukemia cutis. Quality assurance methods and custom equipment enabled safe and effective treatment, achieving precise dosimetry for infants.
Area of Science:
- Medical Physics
- Radiation Oncology
- Pediatric Oncology
Background:
- Total skin electron irradiation (TSEI) is a radiotherapy technique primarily used for adult mycosis fungoides.
- Leukemia cutis isolated to the skin in infants is a rare condition necessitating TSEI.
- This study details the adaptation of TSEI for pediatric patients.
Purpose of the Study:
- To describe the commissioning and quality assurance (QA) for a modified TSEI technique in pediatric patients.
- To ensure accurate and safe radiation delivery for infants with skin-localized leukemia.
- To present the successful clinical implementation of pediatric TSEI.
Main Methods:
- Commissioning involved absolute calibration, dosimetric parameter measurement, and phantom verification using film and OSL dosimeters.
- A modified Stanford technique with a rotating harness system was developed for positioning sedated pediatric patients.
- A comprehensive pediatric TSEI QA program was established, including daily checks and in-vivo dosimetry.
Main Results:
- Dosimetric evaluation in a pediatric phantom confirmed a 2 cm electron penetration depth with surface maximum dose.
- The skin to calibration point dose factor (B-factor) was found to be larger in pediatric TSEI compared to adults.
- In-vivo dosimetry in two infants demonstrated treatment delivery within ±10% of the prescribed dose.
Conclusions:
- Adapting TSEI for pediatric leukemia cutis is feasible using a modified total body irradiation stand.
- The developed commissioning and QA protocols support the clinical implementation of pediatric TSEI.
- Successful treatment of two pediatric patients validates the modified TSEI technique.