Modulation transfer function of a toric intraocular lens: evaluation of the changes produced by rotation and tilt
Adelina Felipe1, José M Artigas, Amparo Díez-Ajenjo
1Fundación Oftalmológica del Mediterráneo and Departamento de Optica, Faculdad de Física, Universidad de Valencia, Valencia, Spain. Adelina.Felipe@uv.es
Purpose:
To evaluate the changes in optical quality when toric intraocular lenses (IOL) are rotated or tilted and to demonstrate that IOL rotation produces an increasing effect of aberrations.
Methods:
Modulation transfer function (MTF) and average modulation were used to analyze the image quality of a toric IOL. The axis of the toric IOL was rotated 5°, 10°, 15°, 20°, 25°, and 30° in successive MTF measurements. The tilt values were 0° to 5°, in increments of 1°, plus a tilt of 15°. Pupil diameters of 3 and 5 mm were used.
Results:
The MTF decay due to aberrations was more sensitive to rotation than tilt. The main decrement in the average modulation, of approximately 50% in both pupils, occurs when the IOL rotates from 0° to 5°. Between 0° and 1° tilt, the average modulation decreases approximately 25% for both pupils; with tilt >1°, the average modulation remains virtually unchanged. The points representing average modulation versus rotation angle are satisfactorily fitted by an exponential function, R=0.98. Average modulation versus tilt angle provides lower correlation degrees, R=0.91 and R=0.79, for 3- and 5-mm pupils, respectively.
Conclusions:
The MTF of the toric IOL decays with rotation and tilt, with greater decrement occurring in rotation from 0° to 5°. An asymptotic value exists in the average modulation decay, meaning that for angles higher than 15°, the average modulation of the toric IOL remains virtually constant.
Related Concept Videos
Mechanism of Ciliary Motion
The cilia are made up of microtubules in a 9+2 arrangement, with nine microtubule doublet ring bundles, surrounding a pair of central singlet microtubule bundles. The doublet microtubule bundles are...
Mechanism of Ciliary Motion
The cilia are made up of microtubules in a 9+2 arrangement, with nine microtubule doublet ring bundles, surrounding a pair of central singlet microtubule bundles. The doublet microtubule bundles are...
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...