IOLMaster versus Manual Keratometry after Photorefractive Keratectomy
Hasan Razmju1, Leila Rezaei, Kobra Nasrollahi
1Isfahan Eye Research Center, Feiz Hospital, Isfahan University of Medical Sciences, Isfahan, Iran.
Journal of Ophthalmic & Vision Research
|March 29, 2012
Summary
Keratometric measurements after photorefractive keratectomy (PRK) using the IOLMaster and a Javal manual keratometer show comparability. However, both methods deviate significantly from the clinical history method (CHM), potentially leading to misleading outcomes.
Area of Science:
- Ophthalmology
- Refractive Surgery
- Ophthalmic Instrumentation
Background:
- Accurate keratometric measurements are crucial for refractive surgery outcomes.
- Photorefractive keratectomy (PRK) alters corneal curvature, necessitating reliable postoperative measurements.
- Comparing different keratometry devices is essential for clinical practice.
Purpose of the Study:
- To compare keratometric measurements obtained by the IOLMaster and a Javal type manual keratometer.
- To evaluate these measurements in eyes that have undergone photorefractive keratectomy (PRK) for myopia.
Main Methods:
- A comparative case series involving patients aged 21-27 years undergoing myopic PRK.
- Preoperative and three-month postoperative keratometry performed using both Javal manual keratometer and IOLMaster.
- Postoperative measurements compared against the clinical history method (CHM).
Main Results:
- Seventy eyes of 35 patients were analyzed, with a mean preoperative spherical equivalent of -4.53 D.
- Postoperative mean measurements: IOLMaster 38.03 D, manual keratometry 43.15 D.
- 95% limits of agreement with CHM: IOLMaster (-5.95 to -0.85), manual keratometry (-1.44 to 4.04).
Conclusions:
- IOLMaster and Javal manual keratometer measurements are comparable to each other after PRK.
- Both devices demonstrate significant deviation from CHM measurements.
- Postoperative keratometry from these devices may yield misleading results.