Related Experiment Video
Updated: May 23, 2026

Continuous Manual Exchange Transfusion for Patients with Sickle Cell Disease: An Efficient Method to Avoid Iron Overload
Published on: March 14, 2017
Clinical evaluation of the Sorin Xtra(R) Autotransfusion System
E P Overdevest1, P W J Lanen, J C M Feron
1Department of Extracorporeal Circulation and Blood Management, Catharina Hospital Eindhoven, Eindhoven, the Netherlands. ed@elive.nl
Abstract:
The performance of the Sorin Xtra® Autotransfusion System (ATS) was studied in 62 patients undergoing coronary artery bypass grafting. Blood was collected intraoperatively and washed using three different wash sets in 4 groups. Both collected and washed blood were analysed for hemoglobin levels and hematocrit, concentrations of proteins, albumin, heparin and plasma free hemoglobin (PFH) were determined, erythrocytes, platelets and leukocytes were counted. Hematocrit measurements of the Xtra® were compared with laboratory measurements to study the accuracy of the Xtra® hematocrit sensor. In addition, the red blood cell recovery rate and elimination rates were calculated to evaluate the clinical performance of the Xtra®. The Xtra® ATS produced a volume of concentrated red blood cells with an average hematocrit from 58% to 63%, depending on the size of the bowl and the chosen default program. In all bowl sizes and programs, the Xtra® Hct-out measurement underestimated the CELL-DYN measurement by approximately 15%. The calculated recovery rates for red blood cells (RBC) in the 4 groups ranged from 86.7% to 91.6%. Elimination rates were calculated in each group for proteins (96.8-99.2%), albumin (96.4-98.7%), plasma free hemoglobin (83.6-91.2%), heparin (98.8-99.9%), platelets (82.4-94.3%) and white blood cells (28.6-42.3%). The Xtra® ATS can be appealing for its performance by producing high hematocrit levels in the washed RBC volume, while keeping RBC recovery rate at the same high level (≈ 90%) as in its predecessor, the Electa® Autotransfusion System.
Related Concept Videos
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy