Sample Drift Correction Following 4D Confocal Time-lapse Imaging
Published on: April 12, 2014
Rapid measurement and correction of phase errors from B0 eddy currents: impact on image quality for non-Cartesian
Ethan K Brodsky1, Jessica L Klaers, Alexey A Samsonov
1Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA. brodskye@cae.wisc.edu
Abstract:
Non-Cartesian imaging sequences and navigational methods can be more sensitive to scanner imperfections that have little impact on conventional clinical sequences, an issue which has repeatedly complicated the commercialization of these techniques by frustrating transitions to multicenter evaluations. One such imperfection is phase errors caused by resonant frequency shifts from eddy currents induced in the cryostat by time-varying gradients, a phenomenon known as B(0) eddy currents. These phase errors can have a substantial impact on sequences that use ramp sampling, bipolar gradients, and readouts at varying azimuthal angles. We present a method for measuring and correcting phase errors from B(0) eddy currents and examine the results on two different scanner models. This technique yields significant improvements in image quality for high-resolution joint imaging on certain scanners. This result suggests that correcting short-time B(0) eddy currents that do not affect conventional clinical sequences may simplify the adoption of non-Cartesian methods.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...