Growth detection failures by the nonradiometric Bactec MGIT 960 mycobacterial culture system
Jeremy A Peña1, Mary Jane Ferraro, Colleen G Hoffman
1Department of Pathology, Massachusetts General Hospital and Harvard Medical School, Boston, Massachusetts, USA.
Abstract:
Mycobacterial growth in liquid culture can go undetected by automated, nonradiometric growth detection systems. In our laboratory, instrument-negative tubes from the Bactec MGIT 960 system are inspected visually for clumps suggestive of mycobacterial growth, which (if present) are examined by acid-fast smear analysis. A 3-year review demonstrated that ∼1% of instrument-negative MGIT cultures contained mycobacterial growth and that 10% of all cultures yielding mycobacteria were instrument negative. Isolates from instrument-negative MGIT cultures included both tuberculous and nontuberculous mycobacteria.