Related Experiment Video
Updated: May 22, 2026

A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
Published on: June 2, 2015
Injection of gadolinium contrast through pediatric central venous catheters: a safety study
John M Moriarty1, Geoffrey L Kung, Yanerys Ramos
1Department of Radiological Sciences, David Geffen School of Medicine, University of California, Los Angeles, CA 90095-7206, USA. jmoriarty@mednet.ucla.edu
Background:
Catheter rupture during CT angiography has prompted policies prohibiting the use of electronic injectors with peripherally inserted central venous catheters (PICCs) not only for CT but also for MRI. Consequently, many institutions mandate hand injection for MR angiography, limiting precision of infusion rates and durations of delivery.
Objective:
To determine whether electronic injection of gadolinium-based contrast media through a range of small-caliber, single-lumen PICCs would be safe without risk of catheter rupture over the range of clinical protocols and determine whether programmed flow rates and volumes were realized when using PICCs for contrast delivery.
Materials And Methods:
Experiments were performed and recorded using the Medrad Spectris Solaris EP MR Injection System. PICC sizes, contrast media and flow rates were based on common institutional protocols.
Results:
No catheters were damaged during any experiments. Mean difference between programmed and delivered volume was 0.07 ± 0.10 mL for all experiments. Reduced flow rates and prolonged injection durations were observed when the injector's pressure-limiting algorithm was triggered, only in protocols outside the clinical range.
Conclusion:
PICCs commonly used in children can withstand in vitro power injection of gadolinium-based contrast media at protocols significantly above clinical levels.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview