Organ-based dose current modulation and thyroid shields: techniques of radiation dose reduction for neck CT
Jenny K Hoang1, Terry T Yoshizumi, Kingshuk Roy Choudhury
1Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA. jennykh@gmail.com
Objective:
The purpose of this study was to assess the difference in absorbed organ dose and image quality for MDCT neck protocols using automatic tube current modulation alone compared with organ-based dose modulation and in-plane thyroid bismuth shielding.
Materials And Methods:
An anthropomorphic female phantom with metal oxide semiconductor field effect transistor (MOSFET) detectors was scanned on a 64-MDCT scanner. The protocols included a reference neck CT protocol using automatic tube current modulation and three modified protocols: organ-based dose modulation, automatic tube current modulation with thyroid shield, and organ-based dose modulation with thyroid shield. Image noise was evaluated quantitatively with the SD of the attenuation value, and subjectively by two neuroradiologists.
Results:
Organ-based dose modulation, automatic tube current modulation with thyroid shield, and organ-based dose modulation with thyroid shield protocols reduced the thyroid dose by 28%, 33%, and 45%, respectively, compared with the use of automatic tube current modulation alone (p ≤ 0.005). Organ-based dose modulation also reduced the radiation dose to the ocular lens (33-47%) compared with the use of automatic tube current modulation (p ≤ 0.04). There was no significant difference in measured noise and subjective image quality between the protocols.
Conclusion:
Both organ-based dose modulation and thyroid shields significantly reduce the thyroid organ dose without degradation of subjective image quality compared with automatic tube current modulation. Organ-based dose modulation has the additional benefit of dose reduction to the ocular lens.
Related Concept Videos
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...