Enhancing pedicle screw fixation in the aging spine with a novel bioactive bone cement: an in vitro biomechanical
Qingan Zhu1, Stephen Kingwell, Zhaoyang Li
1Department of Orthopaedics, University of British Columbia and Vancouver Coastal Health Research Institute, Vancouver, British Columbia, Canada.
Study Design:
A paired biomechanical study of pedicle screws augmented with bone cement in a human cadaveric and osteoporotic lumbar spine model. OBJECTIVES.: To evaluate immediate strength and stiffness of pedicle screw fixation augmented with a novel bioactive bone cement in an osteoporotic spine model and compare it with polymethylmethacrylate (PMMA) cement.
Summary Of Background Data:
A novel bioactive bone cement, containing nanoscale particles of strontium and hydroxyapatite (Sr-HA), can promote new bone formation and osteointegration and provides a promising reinforcement to the osteoporotic spine. Its immediate mechanical performance in augmenting pedicle screw fixation has not been evaluated.
Methods:
Two pedicle screws augmented with Sr-HA and PMMA cement were applied to each of 10 isolated cadaveric L3 vertebrae. Each screw was subjected to a toggling test and screw kinematics were calculated. The pedicle screw was subjected to a pullout test until failure. Finally, the screw coverage with cement was measured on computed tomographic images.
Results:
Screw translations in the toggling test were consistently larger in the Sr-HA group than in the PMMA group (1.4 ± 1.2 mm vs. 1.0 ± 1.1 mm at 1000 cycles). The rotation center was located closer to the screw tip in the Sr-HA group (19% of screw length) than in the PMMA group (37%). The only kinematic difference between Sr-HA and PMMA cements was the screw rotation at 1000 cycles (1.5° ± 0.9° vs. 1.3° ± 0.6°; P = 0.0026). All motion parameters increased significantly with more loading cycles. The pullout force was higher in the PMMA group than the Sr-HA group (1.40 ± 0.63 kN vs. 0.93 ± 0.70 kN), and this difference was marginally significant (P = 0.051). Sr-HA cement covered more of the screw length than PMMA cement (79 ± 19% vs. 43 ± 19%) (P = 0.036).
Conclusion:
This paired-design study identified some subtle but mostly nonsignificant differences in immediate biomechanical fixation of pedicle screws augmented with the Sr-HA cement compared with the PMMA cement.