Biomechanical properties of fixed-angle volar distal radius plates under dynamic loading
William J Dahl1, Paul F Nassab, Kraig M Burgess
1Department of Orthopaedic Surgery, University of Michigan, Ann Arbor, MI 48103-5827, USA.
Purpose:
To evaluate and compare the biomechanical properties of 8 different locked fixed-angle volar distal radius plates under conditions designed to reflect forces seen in early fracture healing and postoperative rehabilitation.
Methods:
We evaluated the Acumed Acu-Loc (Acumed, Hillsboro, OR), Hand Innovations DVR (Hand Innovations, Miami, FL), SBi SCS volar distal radial plate (Small Bone Innovations, Morrisville, PA), Synthes volar distal radius plate and EA extra-articular volar distal radius plate (Synthes, Paoli, PA), Stryker Matrix-SmartLock (Stryker Leibinger, Kalamazoo, MI), Wright Medical Technology Locon VLS (Wright Medical Technology, Arlington, TN), and Zimmer periarticular distal radius locking plate (Zimmer, Warsaw, IN). After affixing each plate to a synthetic corticocancellous radius, we created a standardized dorsal wedge osteotomy. Each construct had cyclic loading of 100 N, 200 N, and 300 N for a total of 6000 cycles. Outcomes, including load deformation curves, displacement, and ultimate yield strengths, were collected for each construct.
Results:
The Wright plate was significantly stiffer at the 100 N load than the Zimmer plate and was stiffer at the 300 N load than 4 other plates. The Zimmer and Hand Innovations plates had the highest yield strengths and significantly higher yield strengths than the Wright, SBi, Stryker, and Synthes EA plates.
Conclusions:
Given the biomechanical properties of the plates tested, in light of the loads transmitted across the native wrist, all plate constructs met the anticipated demands. It seems clear that fracture configuration, screw placement, cost, and surgeon familiarity with instrumentation should take priority in selecting a plating system for distal radius fracture treatment.
Clinical Relevance:
This study provides further information to surgeons regarding the relative strengths of different plate options for the treatment of distal radius fractures.
Related Concept Videos
Normal Strain under Axial Loading
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Impact Loading
In cases of elastic deformation,...