Clinical evaluation of the Oculus Keratograph
N Best1, L Drury, J S Wolffsohn
1Specsavers, 41 High Row, Darlington, Co Durham DL3 7QW, UK.
Summary
The Oculus Keratograph measures corneal curvature and non-invasive tear break-up time (NITBUT) less accurately than other devices. Calibration adjustments are needed for clinical use.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Optics
Background:
- Accurate measurement of corneal curvature and tear film stability is crucial for diagnosing and managing ocular surface diseases.
- The Oculus Keratograph is a device used for assessing these parameters, but its clinical validity and reliability require thorough investigation.
Purpose of the Study:
- To evaluate the validity and reliability of corneal curvature and non-invasive tear break-up time (NITBUT) measurements obtained using the Oculus Keratograph.
- To compare Keratograph measurements with established devices and assess their correlation with patient-reported symptoms.
Main Methods:
- Corneal curvature was measured using the Oculus Keratograph and Nidek ARKT TonorefII in 100 eyes.
- Non-invasive tear break-up time (NITBUT) was assessed objectively with the Keratograph and subjectively with the Keeler Tearscope.
- Repeatability of Keratograph measurements and correlation with the Ocular Surface Disease Index (OSDI) were analyzed.
Main Results:
- The Keratograph reported significantly flatter corneal curvatures compared to the Nidek ARKT TonorefII but demonstrated good repeatability.
- Keratograph NITBUT measurements were significantly lower than Tearscope assessments and showed a decrease on repeat measurements.
- A higher percentage of subjects had NITBUTs below 10 seconds when measured by the Keratograph.
- Tearscope results showed a better correlation with patient symptoms than Keratograph measurements.
Conclusions:
- The Oculus Keratograph requires a calibration offset for accurate comparison with other keratometry devices.
- The current Keratograph software detects very early tear film changes, resulting in lower NITBUT values than subjective methods.
- Software adjustments could improve the clinical utility of the Keratograph's objective tear film assessments.