Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers
Nitin Shah1, Hamsa B Ragaswamy, Kavitha Govindugari
1Department of Anesthesiology, Long Beach Veterans Affairs Healthcare System, Long Beach, CA 90822, USA. nitinshahmd@gmail.com
Study Objective:
To evaluate pulse oximeter performance during motion and induced low perfusion in volunteers.
Design:
Prospective volunteer study.
Setting:
Direct Observation unit.
Subjects:
10 healthy adult volunteers.
Interventions:
Ten volunteers were monitored with three different pulse oximeters while they underwent desaturation to about 75% oxygen saturation (SpO(2)) and performed machine-generated (MG) and volunteer-generated (VG) hand movements with the test hand, keeping the control hand stationary.
Measurements:
SpO(2) and pulse rate readings from the motion (test) and stationary (control) hands were recorded as well as the number of times and the duration that the oximeters connected to the test hands did not report a reading. Sensitivity, specificity, performance index for SpO(2), and pulse rate (PR) were calculated for each pulse oximeter by comparing performance of the test hand with the control hand.
Main Results:
During both MG and VG motion, the Masimo Radical had higher SpO(2) specificity (93% and 97%) than the Nellcor N-600 (67% and 77%) or the Datex-Ohmeda TruSat (83% and 82%). The Masimo Radical also had higher SpO(2) sensitivity (100% and 95%) than the Nellcor N-600 (65% and 50%) or the Datex-Ohmeda TruSat (20% and 15%) during both MG and VG motion. During MG motion, the Masimo Radical had the lowest PR failure rate (0%) compared with the Nellcor N-600 (22.2%) and Datex-Ohmeda TruSat (1.3%). However, during VG motion, the Masimo Radical had the lowest SpO(2) failure rate (0%) of the three devices (Nellcor N-600 16.4% and Datex-Ohmeda TruSat 1.7%). Both the Masimo Radical and the Datex-Ohmeda TruSat had lower PR failure rates (0% and 4.4%) than the Nellcor N-600 (33.9%). There were no significant differences in SpO(2) or PR performance index between the three devices.
Conclusions:
The Masimo Radical had higher SpO(2) sensitivity and specificity than the Nellcor N-600 and Datex-Ohmeda TruSat during conditions of motion and induced low perfusion in this volunteer study.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring pulse