Dichrometer errors resulting from large signals or improper modulator phasing
1Physics Department, East Carolina University, Greenville, NC 27858-4353, USA. sutherlandj@ecu.edu
Abstract:
A single-beam spectrometer equipped with a photoelastic modulator can be configured to measure a number of different parameters useful in characterizing chemical and biochemical materials including natural and magnetic circular dichroism, linear dichroism, natural and magnetic fluorescence-detected circular dichroism, and fluorescence polarization anisotropy as well as total absorption and fluorescence. The derivations of the mathematical expressions used to extract these parameters from ultraviolet, visible, and near-infrared light-induced electronic signals in a dichrometer assume that the dichroic signals are sufficiently small that certain mathematical approximations will not introduce significant errors. This article quantifies errors resulting from these assumptions as a function of the magnitude of the dichroic signals. In the case of linear dichroism, improper modulator programming can result in errors greater than those resulting from the assumption of small signal size, whereas for fluorescence polarization anisotropy, improper modulator phase alone gives incorrect results. Modulator phase can also impact the values of total absorbance recorded simultaneously with linear dichroism and total fluorescence.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
NMR Spectrometers: Resolution and Error Correction
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...