Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
A new approach for atraumatic implant explantation and immediate implant installation
1eduardoanitua@eduardoanitua.com
Objective:
This study reports a new approach for implant extraction in a convenient, fast, and atraumatic way and allowing a new implant placement at same surgical time.
Study Design:
The Biotechnology Institute extraction kit consists of a wrench that allows a 200 N-cm countertorque force, an internal connection extractor, an extractor for external connection, and a set of ratchet handle extension pieces. In this study, a total of 91 implants installed in 42 patients were explanted.
Results:
Extraction torques ranged from 80 to 200 N-cm. In those cases in which implant removal torques exceeded 200 N-cm, a 2-3-mm depth incision was done using a set of atraumatic trephines to avoid too high countertorque forces.
Conclusions:
The possibility of extracting osseointegrated implants opens new doors in oral implantology. The extraction kit allows dentists to re-treat cases and improve the final results, achieving the best outcomes for patients.