A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Rise in ICD shock impedance: lead fracture or death?
Roland X Stroobandt1, Frederic E Van Heuverswyn, Andreas Kucher
1Heart Center, University Hospital, Ghent, Belgium. roland@stroobandt.com
Background:
Remote monitoring allows for interrogation and extensive data retrieval of implantable cardioverter-defibrillators (ICDs). Data on ICD parameters at the time of death and afterwards are limited. The purpose of this retrospective study was to examine the changes in lead impedances of ICDs at the time of death and afterwards.
Methods:
A total of 37 Biotronik (SE & CO. KG, Berlin, Germany) ICDs (20 ICD-cardiac resynchronization therapy, 16 dual-chamber ICDs, and one single-chamber ICD), retrieved after death, were interrogated. Stored intracardiac electrograms were analyzed to determine the cause of death. Impedance trend curves of shock and pacing lead impedances were analyzed and correlated retrospectively with the reported time of death. The influence of cold exposure on lead impedances was tested in three other single-chamber Biotronik ICDs.
Results:
Of 37 patients, the cause of death was due to ventricular tachyarrhythmias in 21 patients. In 12 patients, death was not arrhythmia-related. In four patients, the cause of death could not be determined due to overwriting of the episodes at the time of death. A significant increase of shock and pacing lead impedances was observed in the postmortem days (P < 0.001 for all lead impedances). All lead impedance values increased significantly within the first postmortem day (P < 0.001 for all lead impedances). Cold exposure decreased shock lead impedance but did not affect pacing lead impedance.
Conclusion:
Postmortem analysis of ICDs allows tracking of lead impedance changes, which correlate with the day of death. The rise in postmortem impedances should not be interpreted as contributing to the mode of death.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Secondary Spinal Cord Injury llI: Pathophysiology
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...