Related Experiment Video
Updated: May 21, 2026

Development and Functionalization of Electrolyte-Gated Graphene Field-Effect Transistor for Biomarker Detection
Published on: February 1, 2022
A reusable impedimetric aptasensor for detection of thrombin employing a graphite-epoxy composite electrode
Cristina Ocaña1, Mercè Pacios, Manel del Valle
1Department of Chemistry, Universitat Autònoma de Barcelona, Bellaterra 08193, Spain. cristina.ocana@uab.es
Abstract:
Here, we report the application of a label-free electrochemical aptasensor based on a graphite-epoxy composite electrode for the detection of thrombin; in this work, aptamers were immobilized onto the electrodes surface using wet physical adsorption. The detection principle is based on the changes of the interfacial properties of the electrode; these were probed in the presence of the reversible redox couple [Fe(CN)(6)](3-)/[Fe(CN)(6)](4-) using impedance measurements. The electrode surface was partially blocked due to formation of aptamer-thrombin complex, resulting in an increase of the interfacial electron-transfer resistance detected by Electrochemical Impedance Spectroscopy (EIS). The aptasensor showed a linear response for thrombin in the range of 7.5 pM to 75 pM and a detection limit of 4.5 pM. The aptasensor was regenerated by breaking the complex formed between the aptamer and thrombin using 2.0 M NaCl solution at 42 °C, showing its operation for different cycles. The interference response caused by main proteins in serum has been characterized.