1The Center for Medical Device Evaluation, SFDA, Beijing 100044. yangxd@cmde.org.cn
Objective:
To Investigate necessity of safety evaluation, research ideas, evaluation methods of DEHP in PVC medical devices.
Methods:
The ideas of The United States, Japan and the European Union on safety evaluation of DEHP in PVC medical devices was reviewed and sorted, regarding to currently requirements for PVC medical devices in our country, the research ideas and methods of DEHP in PVC medical devices was explored.
Results:
DEHP released from High-risk PVC medical devices may exceed human tolerance intake values and thus potentially be harmful to human health. So it is necessary for production enterprise to assess safety of DEHP in PVC medical device. The assessment can be done by material control and detecting release of DEHP from PVC medical devices.
Conclusion:
In order to assess safety of DEHP in PVC medical device, production enterprise can firstly assess materials according to national standard. Secondly, production enterprise can detect release of DEHP from PVC medical devices simulating clinical application. By comparing release of DEHP from PVC medical devices and TI, safety of DEHP in PVC medical device can be evaluated.
More Related Videos
10:15The Use of the Ex Vivo Chandler Loop Apparatus to Assess the Biocompatibility of Modified Polymeric Blood Conduits
Published on: August 20, 2014
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
PPE Use in Healthcare Settings II: Doffing
PPE Use in Healthcare Settings I: Donning
Bioplastics
Personal Protective Equipment