Related Experiment Video
Updated: May 21, 2026

Phase Contrast and Differential Interference Contrast (DIC) Microscopy
Published on: August 6, 2008
Convenient contrast enhancement by a hole-free phase plate
Marek Malac1, Marco Beleggia, Masahiro Kawasaki
1National Institute for Nanotechnology, 11421 Saskatchewan Drive, Edmonton, Canada T6G2M9. mmalac@ualberta.ca
Abstract:
Decrease of the irradiation dose needed to obtain a desired signal-to-noise ratio can be achieved by Zernike phase-plate imaging. Here we present results on a hole-free phase plate (HFPP) design that uses the incident electron beam to define the center of the plate, thereby eliminating the need for high precision alignment and with advantages in terms of ease of fabrication. The Zernike-like phase shift is provided by a charge distribution induced by the primary beam, rather than by a hole in the film. Compared to bright-field Fresnel-mode imaging, the hole-free phase plate (HFPP) results in two- to four-fold increase in contrast, leading to a corresponding decrease in the irradiation dose required to obtain a desired signal-to-noise ratio. A local potential distribution, developed due to electron beam-induced secondary-electron emission, is the most likely mechanism responsible for the contrast-transfer properties of the HFPP.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Confocal Fluorescence Microscopy