Development of simple and rapid LC-MS/MS method for determination of celecoxib in human plasma and its application to
Mi-Sun Park1, Wang-Seob Shim, Sung-Vin Yim
1College of Pharmacy, Kyung Hee University, Hoegi-Dong, Dongdaemun-Gu, Seoul 130-701, Republic of Korea.
Abstract:
A suitable liquid chromatography tandem mass spectrometry (LC-MS/MS) method to determine celecoxib in human plasma is needed for bioequivalence and pharmacokinetic studies of celecoxib preparations. The present study describes a simple, rapid, reproducible, and reliable LC-MS/MS method to determine celecoxib concentrations in human plasma. After one-step liquid-liquid extraction (LLE) using methyl tert-butyl ether (MTBE), celecoxib and atorvastatin (internal standard, IS) were eluted on a Luna HILIC column with an isocratic mobile phase, consisting of 10mM ammonium formate buffer (adjusted to pH 3.0 with formic acid):methanol (5:95, v/v) at a flow rate of 0.2 mL/min. The achieved lower limit of quantitation (LLOQ) was 10 ng/mL (S/N>10) and the standard calibration curve for celecoxib was linear (correlation coefficients were >0.9995) over the studied concentration range (10-2000 ng/mL). The inter- and intra-assay coefficients of variation ranged from 1.15% to 4.93% and 1.08% to 7.81%, respectively. The chromatographic run time for each plasma sample was <2 min. The developed method was successfully applied to a bioequivalence study of celecoxib in healthy Korean male volunteers.