Fluoroscopic screening of asymptomatic patients implanted with the recalled Riata lead family
Jeffrey Liu1, Rohit Rattan, Evan Adelstein
1Heart and Vascular Institute, University of Pittsburgh Medical Center, Pittsburgh, PA 15213, USA.
Background:
The Food and Drug Administration recently issued a class I recall of the St. Jude Medical Riata implantable cardioverter-defibrillator lead presumably because of increased risk of electric failure and mechanical separation via inside-out abrasion. We sought to examine the incidence and time dependence of inside-out abrasion in asymptomatic patients implanted with the Riata lead.
Methods And Results:
Asymptomatic patients implanted with the Riata lead at our institution were offered voluntary fluoroscopic screening in 3 views. Electric testing of the Riata lead with provocative isometric muscle contraction was performed at the time of fluoroscopic screening. Of the 245 patients undergoing fluoroscopic screening, 53 (21.6%) patients showed clear evidence of lead separation. Of these externalized leads, 0%, 13%, and 26% had a dwell time of <3 years, 3 to 5 years, and >5 years, respectively (P=0.037). Externalized leads had a significantly pronounced decrease in R-wave amplitude (-1.7±2.9 mV versus +0.35±2.5 mV; P<0.001), and more patients with externalized leads had ≥25% decrease in R-wave amplitude from baseline (28.0% versus 8.1%; P=0.018). One patient with externalization exhibited new noise on near-field electrogram.
Conclusions:
The Riata lead exhibits time-dependent high rates of cable externalization exceeding 20% at >5 years of dwell time. Externalized leads are associated with a more pronounced decrease in R-wave amplitude, which may be an early marker of future electric failure. The use of fluoroscopic and electric screening of asymptomatic patients with the Riata lead remains controversial in the management of patients affected by the recent Food and Drug Administration recall.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI