Related Experiment Video
Updated: May 20, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
In vivo evaluation of the Raypex 5 by using different irrigants
Susana Gomes1, Rogelio Oliver, Carlos Macouzet
1Department of Restorative Dentistry and Endodontics, Universitat Internacional de Catalunya, Barcelona, Spain.
Introduction:
The Raypex 5 is a fourth-generation electronic apex locator for which the accuracy in the presence of chlorhexidine (CHX) and ethylenediaminetetraacetic acid (EDTA) has not yet been tested in vivo. The aim of this in vivo study was to evaluate the performance of the Raypex 5 electronic apex locator in the presence of different irrigant solutions: 2.5% sodium hypochlorite (NaOCl), 2% CHX, and 17% EDTA.
Methods:
Thirty-four single-rooted human teeth that were scheduled for extraction were selected for the study. Measurements were performed with the Raypex 5 in the presence of different irrigant solutions: 2.5% NaOCl, 2% CHX, and 17% EDTA. After the teeth were extracted, a #10 K-file was used to determine the reference working length (RWL), which was established to 0.5 mm from the major foramen. The measurements of WL obtained with the different irrigants were compared by analysis of variance. Significance was set at P < .05.
Results:
No significant differences were found among the experimental groups (P = .18). The mean distance from the RWL to the file tip was -0.26 ± 1.14 mm when 17% EDTA was used, -0.03 ± 0.92 mm for 2% CHX, and 0.22 ± 0.93 mm for 2.5% NaOCl.
Conclusions:
The Raypex 5 performed equally well irrespective of the irrigant used.
Related Concept Videos
In Vitro Drug Dissolution: Alternative Methods
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...