Low-dose protocol for head CT in evaluation of hydrocephalus in children
Katarzyna Jończyk-Potoczna1, Marta Frankiewicz, Małgorzata Warzywoda
1Department of Pediatric Radiology of the Chair of Radiology, Poznań University of Medical Sciences, Poznań, Poland.
Background:
A suspicion of ventriculo-peritoneal shunt failure is classified as the most common indication for CT in children with hydrocephalus. The main target of the study was to evaluate the diagnostic value of a low-dose protocol and to compare a total DLP received by patients in compared protocols.
Material/Methods:
Our retrospective analysis included 256 examinations performed in patients aged from 1 month to 18 years, with body mass ranging from 3 to 100 kg. The examinations were conducted in the years 2009-2011. A total number of 128 examinations were performed on the basis of the low-dose protocol and 128 according to a standard protocol using the Siemens SOMATOM Definition AS 128-slice scanner. RESULTS/CONLUSIONS: The analysis showed a full value of the diagnostic low-dose protocol with a simultaneous decrease of the total dose of DLP to the average of 40%. Application protocol with lower mAs in assessing the causes of ventriculo-peritoneal shunt failure in children with hydrocephalus is coherent with the valid principles of radiation protection in pediatrics and reduces the total DLP while maintaining a very good diagnostic value.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...