Abstract:
Ionization damage (radiolysis) and knock-on displacement are compared in terms of scattering cross section and stopping power, for thin organic specimens exposed to the electrons in a TEM. Based on stopping power, which includes secondary processes, radiolysis is found to be predominant for all incident energies (10-300 keV), even in materials containing hydrogen. For conducting inorganic specimens, knock-on displacement is the only damage mechanism but an electron dose exceeding 1000 C cm(-2) is usually required. Ways of experimentally determining the damage mechanism (with a view to minimizing damage) are discussed.