Dual-scan acquisition for accelerated continuous-wave EPR oximetry

J Palmer1, L C Potter, D H Johnson

  • 1Department of Electrical and Computer Engineering, The Ohio State University, Columbus, OH 43210, USA.

Summary

Dual-scan acquisition improves oxygen sensing precision and reduces spectroscopic acquisition time by 3-4 times compared to single-scan methods. This technique enhances linewidth determination for precise partial pressure of oxygen (pO(2)) measurements.