Related Experiment Video
Updated: May 20, 2026

Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
Published on: January 5, 2024
Variance analysis of x-ray CT sinograms in the presence of electronic noise background
Jianhua Ma1, Zhengrong Liang, Yi Fan
1Department of Radiology, State University of New York, Stony Brook, NY, USA.
Purpose:
Low-dose x-ray computed tomography (CT) is clinically desired. Accurate noise modeling is a fundamental issue for low-dose CT image reconstruction via statistics-based sinogram restoration or statistical iterative image reconstruction. In this paper, the authors analyzed the statistical moments of low-dose CT data in the presence of electronic noise background.
Methods:
The authors first studied the statistical moment properties of detected signals in CT transmission domain, where the noise of detected signals is considered as quanta fluctuation upon electronic noise background. Then the authors derived, via the Taylor expansion, a new formula for the mean-variance relationship of the detected signals in CT sinogram domain, wherein the image formation becomes a linear operation between the sinogram data and the unknown image, rather than a nonlinear operation in the CT transmission domain. To get insight into the derived new formula by experiments, an anthropomorphic torso phantom was scanned repeatedly by a commercial CT scanner at five different mAs levels from 100 down to 17.
Results:
The results demonstrated that the electronic noise background is significant when low-mAs (or low-dose) scan is performed.
Conclusions:
The influence of the electronic noise background should be considered in low-dose CT imaging.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Random Error
Interpreting X̄ Charts
An x̄ chart plots the values of individual measurements over time against control limits calculated from historical data. The central line represents the process mean,...
Variance
Imaging Studies III: Computed Tomography