Related Experiment Video
Updated: May 20, 2026

Treatment of Facial Deformities using 3D Planning and Printing of Patient-Specific Implants
Published on: May 23, 2020
Three-dimensional mandibular change after sagittal split ramus osteotomy with a semirigid sliding plate system for
Moon-Ho Ghang1, Hyung-Min Kim, Jun-Young You
1Department of Oral and Maxillofacial Surgery, School of Dentistry, Seoul National University, Seoul, Korea.
Objective:
This study proposes a new plate system, a sliding plate, after sagittal split ramus osteotomy (SSRO). The sliding plate was designed to adjust the spatial relationship between the proximal and distal segments during the postoperative period. This plate was used for mandibular setback surgery to minimize forward early relapse.
Study Design:
Twenty-six Korean subjects who had undergone mandibular setback surgery by SSRO were included in this study. Skeletal stability was evaluated with lateral cephalograms and 3-dimensional cone-beam computerized tomography.
Results:
Forward mandibular movement during the postoperative period was ~2 mm. The condyles rotated inward on the axial view after surgery. Postoperatively, the condyle rotated outward on the axial view, inward on the coronal view, and forward on the sagittal view by a statistically significant amount.
Conclusions:
The fixation method with sliding plates after SSRO was stable and convenient for the operators during surgery and postoperative care.