Related Experiment Video
Updated: May 20, 2026

The Application of 1% Methylene Blue Dye As a Single Technique in Breast Cancer Sentinel Node Biopsy
Published on: June 1, 2019
Detection efficiency for measuring 241Am in axillary lymph nodes using different types and sizes of detectors
Megan L Lobaugh1, Henry B Spitz, Samuel E Glover
1University of Cincinnati, Nuclear and Radiological Engineering Program, Cincinnati, OH 45221-0072, USA. lobaugml@mail.uc.edu
Abstract:
The detection efficiency and interference susceptibility of four different types of low energy photon detectors, each with a unique geometric arrangement, were compared for direct measurement of Am deposited in the axillary lymph nodes. Although the most efficient detector was a single large 23,226 mm square phoswich detector, it was also the most susceptible to confounding depositions from activity deposited in adjacent organs. The array of two 2,800 mm high purity germanium detectors exhibited the highest efficiency per unit detector area with some resistance to confounding from activity deposited in the lungs. The array of two 4,560 mm NaI(Tl) detectors was the least susceptible to confounding and nearly as efficient per square millimeter as the high purity germanium detector array. Thus, selection of a detector system for in vivo measurement of activity deposited in the axillary lymph nodes should consider whether there is a likelihood for activity deposited in other organs, such as the lungs, skeleton, or liver, to create an interference that will confound the measurement result.
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Gas Chromatography: Types of Detectors-II
Biological Effects of Radiation
High-Performance Liquid Chromatography: Types of Detectors
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...