Related Experiment Video
Updated: May 19, 2026

Rapid Homogeneous Detection of Biological Assays Using Magnetic Modulation Biosensing System
Published on: June 13, 2010
Immunosensor based on magnetic relaxation switch and biotin-streptavidin system for the detection of Kanamycin in
Yi Ping Chen1, Ming qiang Zou, Cai Qi
1Chinese Academy of Inspection and Quarantine, Beijing 100123, China.
Abstract:
A rapid, sensitive, and simple immunosensor was developed for the detection of Kanamycin (KM) in milk. This immunosensor is based on magnetic relaxation switch (MRS) assay and biotin-streptavidin system (B-SA system). The target analyte (KM) competed with those on the surface of the superparamagnetic iron oxide (SPIO) nanoparticles and hence affected the formation of SPIO aggregates. The dispersed and aggregated states of SPIO can modulate the spin-spin relaxation time (T(2)) of the neighboring water molecule. T(2) was then changed as an effect of the target analyte. The B-SA system was used to amplify the SPIO binding, thus enhance the sensitivity. The detection working was 1.5 to 25.2ng mL(-1) and limit of detection (LOD) was determined to be 0.1ng mL(-1). The LOD of the immunosensor decreased tenfold, and its analysis time (45min) was much shorter than that of enzyme-linked immunosorbent assay (6h to 8h). The average recoveries of the KM at various spiking levels ranged from 80.2% to 85.6% with a relative standard deviation (RSD) below 4.0%. The results showed that the MRS immunosensor was a promising platform for the determination of small molecular residues because of its high sensitivity, specificity, homogeneity, and speed.