Related Experiment Video
Updated: May 19, 2026

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Subject-specific models of susceptibility-induced B0 field variations in breast MRI
Caroline D Jordan1, Bruce L Daniel, Kevin M Koch
1Radiology, Stanford University, Stanford, California, USA. cjordan@stanford.edu
Purpose:
To rapidly calculate and validate subject-specific field maps based on the three-dimensional shape of the bilateral breast volume.
Materials And Methods:
Ten healthy female volunteers were scanned at 3 Tesla using a multi-echo sequence that provides water, fat, in-phase, out-of-phase, and field map images. A shape-specific binary mask was automatically generated to calculate a computed field map using a dipole field model. The measured and computed field maps were compared by visualizing the spatial distribution of the difference field map, the mean absolute error, and the 80% distribution widths of frequency histograms.
Results:
The 10 computed field maps had a mean absolute error of 38 Hz (0.29 ppm) compared with the measured field maps. The average 80% distribution widths for the histograms of all of the computed, measured, and difference field maps are 205 Hz, 233 Hz, and 120 Hz, respectively.
Conclusion:
The computed field maps had substantial overall agreement with the measured field maps, indicating that breast MRI field maps can be computed based on the air-tissue interfaces. These estimates may provide a predictive model for field variations and thus have the potential to improve applications in breast MRI.