Related Experiment Video
Updated: May 19, 2026

In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
The effect of axial length on the variability of Stratus optical coherence tomography
Jeong Hun Bae1, So Young Han, Hyunjoong Kim
1Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Korea.
Purpose:
To evaluate the effect of axial length on the variability of retinal nerve fiber layer (RNFL) thickness measurements using the Stratus optical coherence tomography (OCT) in normal and glaucomatous eyes.
Methods:
We measured the RNFL thickness in 474 subjects using the Stratus OCT twice during the same day. Axial length was measured with the IOLMaster, and refractive error was the absolute value of the spherical equivalent measured with an auto ref-keratometer. Standard deviation in overall mean RNFL thickness was used as the dependent variable to identify significant correlations.
Results:
Long axial length affected the variability in the RNFL thickness value by stratus OCT at the temporal quadrant (p = 0.006) and clock-hour sector 9 (p = 0.001). Refractive error also affected the variability of the RNFL thickness value by stratus OCT at the temporal quadrant (p = 0.025) and clock-hour sector 9 (p = 0.024).
Conclusions:
It is clinically significant that longer axial length demonstrates greater variability in temporal area as detected by OCT, a measurement which correlates with the preferably damaged position in the myopic glaucoma eye.
Related Concept Videos
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Normal Strain under Axial Loading
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Imaging Studies III: Computed Tomography