End tidal carbon dioxide is as reliable as transcutaneous monitoring in ventilated postsurgical neonates
David Gerald Tingay1, Kwok Sean Mun, Elizabeth Jean Perkins
1Department of Neonatology, Royal Children's Hospital, Flemington Road, Parkville, Melbourne, Victoria 3052, Australia. david.tingay@rch.org.au
Objectives:
To compare the agreement, precision and repeatability of end tidal carbon dioxide (EtCO2 ) and transcutaneous carbon dioxide (TcCO2 ) with partial pressure of arterial CO(2) ( PaCO2) in postoperative neonates.
Patients:
Fifty mechanically ventilated neonates without lung disease, and with no contraindications for either TcCO2 or EtCO2 monitoring.
Interventions:
Paired TcCO2 and EtCO2 values were recorded with three consecutive measurements within the first 48 h of surgery.
Main Outcome Measures:
EtCO2, TcCO2 and PaCO2 triplets were compared using Bland-Altman plots.
Results:
One hundred thirty-two triplet measures of CO(2) were recorded with mean PaCO2 43.5 (7.3) mm Hg, EtCO2 38.8 (6.4) mm Hg and 43.8 (8.8) mm Hg (p<0.0001 for EtCO2 against PaCO2; paired t test). The PaCO2 - EtCO2 bias±2SD was 4.1±9.0 mm Hg and -0.8±13.0 mm Hg for PaCO2 - TcCO2. 56.1% of EtCO2, and 60.6% of TcCO2 values were within ±5 mm Hg of paired PaCO2.
Conclusions:
In postoperative neonates, EtCO2 and TcCO2 demonstrated a clinically acceptable agreement with PaCO2.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...