Related Experiment Video
Updated: May 19, 2026

Scleral Cross-linking Using Riboflavin and Ultraviolet-A Radiation for Prevention of Axial Myopia in a Rabbit Model
Published on: April 3, 2016
Evaluation of a new single-piece 4% water content hydrophobic acrylic intraocular lens in the rabbit model
Andrew Ollerton1, Liliana Werner, Stanley R Fuller
1Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah 84132, USA.
Purpose:
To evaluate uveal and capsular biocompatibility of a 1-piece intraocular lens (IOL) manufactured from a new hydrophobic acrylic material that incorporates a barrier step at the optic-haptic junctions.
Setting:
John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
Design:
Experimental study.
Methods:
The study IOL (Eternity-Uni W-60) was implanted in the right eyes of 5 New Zealand rabbits and the control IOL (Acrysof SN60WF) in the left eyes. Slitlamp examination was performed 1 through 4 weeks postoperatively. After death, the globes were enucleated and fixed in formalin. Capsular bag opacification scoring (Miyake-Apple view) was then performed followed by complete histopathology.
Results:
At the 4-week examination, the mean posterior capsule opacification (PCO) score was 1.5 ± 1.0 (SD) in the study group and 2.2 ± 1.09 in the control group (P=.02). Anterior capsule opacification (ACO) was not present in the study eyes and was mild in the control eyes. On gross examination, the mean central PCO score was 0.9 ± 0.65 in the study group and 1.7 ± 1.20 in the control group (P=.07); the mean peripheral PCO score was 1.3 ± 0.67 and 2.4 ± 1.14 (P=.01) and the mean Soemmerring ring score was 3.8 ± 0.44 and 4.2 ± 1.09, respectively (P=.47). Histopathology confirmed that both IOLs were equally tolerated by the rabbit eyes.
Conclusions:
In this rabbit study, the new hydrophobic acrylic material was biocompatible. The barrier step incorporated to the optic-haptic junctions has the potential to enhance PCO prevention.