In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Evaluation of valvular insufficiency and shunts with parallel-imaging compressed-sensing 4D phase-contrast MR imaging
Albert Hsiao1, Michael Lustig, Marcus T Alley
1Department of Radiology, Stanford University School of Medicine, 725 Welch Rd, Room 1679, MC 5913, Stanford, CA 94305-5654, USA. alhsiao@stanford.edu
Purpose:
To assess the potential of compressed-sensing parallel-imaging four-dimensional (4D) phase-contrast magnetic resonance (MR) imaging and specialized imaging software in the evaluation of valvular insufficiency and intracardiac shunts in patients with congenital heart disease.
Materials And Methods:
Institutional review board approval was obtained for this HIPAA-compliant study. Thirty-four consecutive retrospectively identified patients in whom a compressed-sensing parallel-imaging 4D phase-contrast sequence was performed as part of routine clinical cardiac MR imaging between March 2010 and August 2011 and who had undergone echocardiography were included. Multiplanar, volume-rendered, and stereoscopic three-dimensional velocity-fusion visualization algorithms were developed and implemented in Java and OpenGL. Two radiologists independently reviewed 4D phase-contrast studies for each of 34 patients (mean age, 6 years; age range, 10 months to 21 years) and tabulated visible shunts and valvular regurgitation. These results were compared with color Doppler echocardiographic and cardiac MR imaging reports, which were generated without 4D phase-contrast visualization. Cohen κ statistics were computed to assess interobserver agreement and agreement with echocardiographic results.
Results:
The 4D phase-contrast acquisitions were performed, on average, in less than 10 minutes. Among 123 valves seen in 34 4D phase-contrast studies, 29 regurgitant valves were identified, with good agreement between observers (k=0.85). There was also good agreement with the presence of at least mild regurgitation at echocardiography (observer 1, κ=0.76; observer 2, κ=0.77) with high sensitivity (observer 1, 75%; observer 2, 82%) and specificity (observer 1, 97%; observer 2, 95%) relative to the reference standard. Eight intracardiac shunts were identified, four of which were not visible with conventional cardiac MR imaging but were detected with echocardiography. No intracardiac shunts were found with echocardiography alone.
Conclusion:
With velocity-fusion visualization, the compressed-sensing parallel-imaging 4D phase-contrast sequence can augment conventional cardiac MR imaging by improving sensitivity for and depiction of hemodynamically significant shunts and valvular regurgitation.