Arc-ZTE: Continuously-Slewed Zero-TE Imaging With Incoherent Temporal Sampling for Near-Silent Dynamic MRI
Shreya Ramachandran1, Tobias C Wood2, Gavin Zhang1
1Department of Electrical Engineering and Computer Sciences, University of California Berkeley, Berkeley, California, USA.
Purpose:
To enable flexible dynamic and DCE MRI with low acoustic noise and high spatiotemporal resolution using zero echo time (ZTE) imaging.
Methods:
This work proposes Arc-ZTE, a technique in which the readout gradients of ZTE are continuously-slewed to form arcs in k-space that are sequentially rotated to yield an incoherent trajectory. These rotations are selected via an optimization procedure that both promotes temporal k-space coverage and penalizes gradient refocusing. Arc-ZTE was evaluated against existing radial ZTE schemes based on metrics for acoustic noise, k-space coverage uniformity, and sampling incoherence. Phantom and in vivo demonstrations were conducted to evaluate image quality at high temporal resolutions.
Results:
Arc-ZTE achieved stable k-space coverage and improved sampling incoherence over a range of temporal resolutions, with respect to radial ZTE schemes. Although increasing arc curvature increased its acoustic noise over standard radial ZTE (by 0.3-14.2 dB), Arc-ZTE remained substantially quieter than conventional 3D sequences by 20-35 dB. In vivo experiments demonstrated that time-resolved Arc-ZTE effectively captured dynamics of respiratory motion and contrast uptake when reconstructed at 0.5 and 1.1 s/frame respectively.
Conclusion:
Arc-ZTE enables flexible, time-resolved, near-silent dynamic MRI through incoherent temporal sampling, which allows data to be retrospectively binned into a flexible choice of temporal resolution without modifying the trajectory design. This work has the potential to significantly improve the success rates of dynamic imaging among neonates, young children, and other sound-sensitive patients.
More Related Videos
10:143D Scanning Technology Bridging Microcircuits and Macroscale Brain Images in 3D Novel Embedding Overlapping Protocol
Published on: May 12, 2019
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography