[The improvement of determination method of carbon disulfide in workplace air with gas chromatography]
Qing-feng Xiao1, Wen-hua Qin, Yan-yan Lu
1Henan Provincial Institute of Occupational Health, Zhengzhou 450052, China.
Objective:
Establishment of determination method of carbon disulfide in charcoal tube with low toxicity solvents desorption-gas chromatography.
Methods:
Four types of solvent with low toxicity are applied respectively as substitution of benzene to desorb the carbon disulfide in samples of charcoal tube. The signal strength and desorption efficiency of the detector are compared by using different solvents.
Results:
Chloroform has been considered as the best alternative solvent of benzene. Carbon disulfide has a good linearity (R = 0.9997) over the concentration of 0 ∼ 54.7 µg/ml, detection limit can reach 0.2 µg/ml. When the sampling volume is 3.0 L, the minimum detectable concentration is 0.07 mg/m(3).
Conclusion:
With the use of chloroform, the health hazard to laboratory personnel and environment pollution as well as the costs of experiments are reduced.
More Related Videos
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...