[Measurement of 16 elements in blood by inductively coupled plasma mass spectrometry]
Hui-ling Li1, Feng-tong Hao, Jing Ma
1Department of Occupational Disease and Toxication, Capital University of Medical Science, Beijing, China.
Objective:
To establish a method of determining more elements in whole blood by inductively coupled plasma mass spectrometry (ICP-MS).
Methods:
The whole blood samples were treated by 5% nitric acid to remove the proteins, then were centrifuged. Sixteen elements (Be, Al, Mn, V, Cr, Co, Ni, As, Mo, Ag, Cd, Sb, Ba,Tl, Pb and U) in the supernatant were directly measured by ICP-MS.
Results:
The detection limits of 16 elements were 0.01 ∼ 6.51 µg/L. The linear correlation coefficient was ≥ 0.999. The relative standard deviations were below 5 %. The recovery rates were 105% ∼ 115%. Seronorm Trace Elements SerumL-1 LOT 0903106 and GBW09139g or GBW09140g were used in the quality control, the detected results corresponded with the standard values.
Conclusion:
ICP-MS technique is a simple, rapid, accurate and reliable method, which can be used to measure several trace elements in whole blood samples.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Mass Spectrometers