In vitro comparison of two different methods of cell washing.
T Smith1, W Riley, D Fitzgerald
1Brigham and Women's Hospital, Boston, Massachusetts 02115, USA. tsmith17@partners.org
Perfusion
|September 1, 2012
Summary
Washing stored red blood cells (RBCs) with an autotransfusion device significantly reduces harmful substances like free hemoglobin and potassium. This method improves the safety of aged RBC products for patient transfusion.
Area of Science:
- Transfusion Medicine
- Blood Banking
- Biomedical Engineering
Background:
- Red blood cell (RBC) storage leads to accumulation of detrimental substances, known as storage lesion.
- These accumulated substances can cause adverse effects in patients receiving transfusions.
- Autotransfusion devices, typically used in surgery, offer a potential method for washing stored RBCs.
Purpose of the Study:
- To compare the efficacy of washing stored RBC units using an AutoLog autotransfusion device versus a standard blood bank cell washer.
- To evaluate the reduction of key analytes associated with RBC storage lesion after washing.
Main Methods:
- Eleven outdated RBC units (>42 days) were divided and washed using either a Medtronic AutoLog device or a Cobe 2991 Cell Processor.
- Key parameters including hematocrit, free hemoglobin, lactate, lactate dehydrogenase (LDH), potassium, glucose, and pH were measured.
- Analyte concentrations were compared between unwashed parent units and washed daughter units.
Main Results:
- Washing with the AutoLog device significantly decreased glucose, lactate, and potassium levels compared to original RBCs.
- Free hemoglobin and LDH levels were significantly lower when using the Medtronic AutoLog device compared to the Cobe 2991 Cell Processor.
- The AutoLog device effectively reduced harmful analytes associated with RBC storage lesion.
Conclusions:
- Autotransfusion devices can effectively wash stored RBC units, mitigating storage lesion.
- Using an autotransfusion device prior to transfusion attenuates increases in glucose, potassium, free hemoglobin, and LDH.
- This washing method offers a promising approach to improve the safety of stored RBC products.