Assessment of on-screen measurements, magnification, and calibration in digital radiography
Robyn L Read1, Colleen G Duncan, Aaron D Wallace
1Aspen Meadow Veterinary Specialists, Longmont, CO 80501, USA. rread@aspenmeadowvet.com
Objective:
To compare calibration methods for digital radiography in terms of measurement accuracy and interobserver variability.
Design:
Prospective study.
Sample:
Digital radiographic images of a 155-mm-long Steinmann pin.
Procedures:
Measurement of pin length on digital radiographs was determined with a 25.4-mm-diameter calibration ball and commercially available software program via 3 calibration methods (ie, no calibration, autocalibration, and manual calibration). Digital radiographs of the calibration ball and pin were obtained with each placed at various vertical heights from the table (7 heights) and horizontal distances from the center of the beam (4 distances). Measurements of pin length on digital radiographs were made by 4 observers who were blinded to the orientation of the calibration ball and pin.
Results:
Pin lengths obtained by each calibration method were significantly different from each other and from the true value. Manual calibration was the most accurate. There was no significant interobserver variability in measurements. There was no significant change in measurements when the calibration ball was moved horizontally, but pin length measurements changed significantly when the ball was moved vertically (away from the table) with an approximate magnification error of 1% per centimeter of distance between the calibration ball and pin.
Conclusions And Clinical Relevance:
For digital radiography, manual calibration is recommended to achieve the most accurate measurements. Ideally, the calibration ball should be placed at the same vertical height as the object to be measured; however, if this cannot be achieved, the magnification error can be expected to be approximately 1% per centimeter of distance.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT