Related Experiment Video
Updated: May 18, 2026

Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
Comparison of posterior capsule opacification between a 1-piece and a 3-piece microincision intraocular lens
Ana Prinz1, Pia Veronika Vecsei-Marlovits, Dietrich Sonderhof
1Department of Ophthalmology, Hietzing Hospital, Wolkersbergenstrasse 1, Wien 1130, Austria.
Objective:
To compare the intensity of posterior capsular opacification (PCO) between a 1-piece and a 3-piece microincision cataract surgery intraocular lens (MICS IOL) in a prospective randomised study.
Methods:
80 eyes of 40 patients with age-related cataract were enrolled in this study. Each patient received a 1-piece MICS IOL (AF-1 NY-60, Hoya, Tokyo, Japan) in one eye and a 3-piece MICS IOL (AF-1 iMICS Y-60H, Tokyo, Hoya) in the other eye. At the 1-year follow-up, the patients were examined at the slit lamp, visual acuity was determined and standardised high-resolution digital retroillumination images were taken for objective quantification of regeneratory PCO using an automated image analysis software (AQUA).
Results:
The mean regeneratory PCO score (1-piece IOL: 0.2, 3-piece IOL 0.3, p=0.7) and the neodymium:yttrium-aluminium-garnet laser capsulotomy rate (two cases in 3-piece IOL group; p=0.5) were comparable low for both IOLs. Capsular folds occurred significantly more often in the 3-piece IOL group (p=0.02).
Conclusions:
Modification of the MICS IOL from a 3-piece to a 1-piece haptic design caused in short term no significant change in PCO amount. Compared with the 3-piece IOL, the 1-piece IOL led to significantly less capsular folds 1 year after surgery.