Prehospital point-of-care testing for troponin: are the results reliable?
Joseph M Venturini1, Christine E Stake, Mark E Cichon
1Stritch School of Medicine, Loyola University of Chicago, Maywood, IL 60153, USA. jventurini@lumc.edu
Background:
Swift assessment of patients presenting with chest pain results in faster treatment and improved outcomes. Allowing ambulance crews to use point-of-care (POC) devices to measure cardiac troponin I levels during transport of patients to the emergency department (ED) may result in earlier diagnosis of acute myocardial infarction, particularly in those patients without ST-segment elevation. The ability of POC devices to measure cardiac troponin I levels reliably in a moving ambulance has not previously been tested. Objective. This study was conducted to determine whether POC devices operated in a moving ambulance reliably duplicate the measurement of cardiac troponin I levels obtained by POC devices in the ED.
Methods:
Blood samples were obtained in the ED and the hospital from patients reporting chest pain or other cardiac complaints. Troponin I assays were then performed in a moving ambulance using two POC devices. The POC devices were placed on flat surfaces in the rear of the ambulance. The ambulance driver was instructed to keep the ambulance moving in traffic while each assay was completed. A variety of routes were taken. Each set of two assays was completed entirely during a single simulated run. The results of the two assays performed in the moving ambulance were then compared with the results of the control assay, which was performed simultaneously in the ED on the same sample.
Results:
Forty-two whole-blood samples underwent troponin I assays in a moving ambulance. Thirteen (30.9%) assays were positive. One (2.4%) was excluded because of cartridge error. Two (4.8%) were excluded because of interfering substance. No significant difference in whole-blood troponin results was found between the assays performed in the moving ambulance and those performed in the ED (intraclass correlation coefficient 0.997; 95% confidence interval 0.994 to 0.998; p < 0.005).
Conclusions:
When used in a moving ambulance, the POC device provided results of cardiac troponin I assays that were highly correlated to the results when the device was used in the ED. The feasibility, practicality, and clinical utility of prehospital use of POC devices must still be assessed. Key words: point-of-care systems; prehospital emergency care; troponin; reliability of results; ambulances; myocardial infarction; chest pain.
More Related Videos
05:58Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Pneumonia III: Complications and Assessment
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Urine Studies II: Urine Culture and Sensitivity Test