Related Experiment Video
Updated: May 18, 2026

A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
Toward truly combined PET/CT imaging using PET detectors and photon counting CT with iterative reconstruction
Christian Thibaudeau1, Philippe Berard, Marc-Andre Tetrault
1Sherbrooke Molecular Imaging Center, Centre hospitalier universitaire de Sherbrooke, Sherbrooke, Québec, Canada.
Purpose:
This paper intends to demonstrate the feasibility of truly combined PET/CT imaging and addresses some of the major challenges raised by this dual modality approach. A method is proposed to retrieve maximum accuracy out of limited resolution computed tomography (CT) scans acquired with positron emission tomography (PET) detectors.
Methods:
A PET/CT simulator was built using the LabPET™ detectors and front-end electronics. Acquisitions of energy-binned data sets were made using this low spatial resolution CT system in photon counting mode. To overcome the limitations of the filtered back-projection technique, an iterative reconstruction library was developed and tested for the counting mode CT. Construction of the system matrix is based on a preregistered raster scan from which the experimental detector response is obtained. PET data were obtained sequentially with CT in a conventional manner.
Results:
A meticulous description of the system geometry and misalignment corrections is imperative and was incorporated into the matrix definition to achieve good image quality. Using this method, no sinogram precorrection or interpolation is necessary and measured projections can be used as raw input data for the iterative reconstruction algorithm. Genuine dual modality PET/CT images of phantoms and animals were obtained for the first time using the same detection platform.
Conclusions:
CT and fused PET/CT images show that LabPET™ detectors can be successfully used as individual X-ray photon counting devices for low-dose CT imaging of the anatomy in a molecular PET imaging context.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...