Related Experiment Video
Updated: May 18, 2026

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
Published on: December 30, 2025
A sub-ppm acetone gas sensor for diabetes detection using 10 nm thick ultrathin InN FETs
Kun-Wei Kao1, Ming-Che Hsu, Yuh-Hwa Chang
1Institute of NanoEngineering and MicroSystems, National Tsing Hua University, Hsinchu 30013, Taiwan. s9835812@m98.nthu.edu.tw
Abstract:
An indium nitride (InN) gas sensor of 10 nm in thickness has achieved detection limit of 0.4 ppm acetone. The sensor has a size of 1 mm by 2.5 mm, while its sensing area is 0.25 mm by 2 mm. Detection of such a low acetone concentration in exhaled breath could enable early diagnosis of diabetes for portable physiological applications. The ultrathin InN epilayer extensively enhances sensing sensitivity due to its strong electron accumulation on roughly 5-10 nm deep layers from the surface. Platinum as catalyst can increase output current signals by 2.5-fold (94 vs. 37.5 μA) as well as reduce response time by 8.4-fold (150 vs. 1,260 s) in comparison with bare InN. More, the effect of 3% oxygen consumption due to breath inhalation and exhalation on 2.4 ppm acetone gas detection was investigated, indicating that such an acetone concentration can be analyzed in air.