Related Experiment Videos
Computer-assisted assessment of the kinetic function of disk-type artificial heart valves
A G Formanek1, W T Sobol, K M Link
1Department of Radiology, Bowman Gray School of Medicine, Wake Forest University, Winston-Salem, North Carolina 27103.
Investigative Radiology
|January 1, 1990
Abstract:
A simple and accurate computer simulation method was developed to evaluate the geometrical parameters of disk-type artificial heart valves from cinefluorographic recordings. In vitro valve models with preset occluder angles were used to test the accuracy of the method. Linear regression analysis produced slope 1.00055 and intercept 0.0402, with r = 0.999. In vivo reproducibility was tested through repeated measurements, taken several weeks apart, of OMS valves in situ. Regression with slope 1.02 and intercept -0.95, r = 0.95 were found.