Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation.

W Karlen1, K Kobayashi, J M Ansermino

  • 1Electrical & Computer Engineering in Medicine Group, Department of Electrical & Computer Engineering, The University of British Columbia, 2332 Main Mall, Vancouver, BC, V6T 1Z4, Canada. walter.karlen@ieee.org

Physiological Measurement
|September 19, 2012
PubMed
Summary

A new algorithm accurately identifies pulse oximetry signals and assesses their quality in real time. This method improves pulse detection and reduces false alarms, aiding users and alarm system design.

Related Concept Videos

Correlation between ECG and Cardiac Cycle01:25

Correlation between ECG and Cardiac Cycle

The electrical signals recorded on an electrocardiogram (ECG) occur before the mechanical processes of contraction and relaxation during the cardiac cycle.
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Pulse amplitude and quality01:17

Pulse amplitude and quality

Pulse amplitude is a crucial indicator of cardiac health because it provides valuable insights into the strength of left ventricular contractions and the overall uniformity of blood circulation within the vasculature. The strength of the pulse is directly related to the force with which the heart contracts and the volume of blood being pumped.
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure01:17

Measurement of Blood Pressure

Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a stethoscope.