Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Liver MRI at 3.0 tesla: comparison of image quality and lesion detectability between single-source conventional and
Jiyoung Hwang1, Young Kon Kim, Min Jung Park
1Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Objective:
To prospectively and intraindividually compare liver magnetic resonance imaging (MRI) using single-source and dual-source parallel radiofrequency (RF) transmissions at 3.0-T for image quality, lesion detectability, and lesion contrast.
Methods:
Ninety-nine patients with 139 liver lesions underwent liver MRI at 3.0-T. Two radiologists performed a consensus review of T2-weighted images (T2WI), heavily T2WI (HT2WI), gadoxetic acid-enhanced hepatobiliary images, and diffusion-weighted imaging using single-source and dual-source RF transmissions with regard to image quality and lesion detectability. Contrast ratios between liver lesions and liver parenchyma were also calculated.
Results:
Image quality was better with dual-source than with single-source at T2WI and HT2WI (P < 0.05), but lesion detectabilities were similar for all sequences. There was no significant difference in mean contrast ratios for all sequences (P > 0.05).
Conclusion:
Dual-source RF transmission provides a better image quality with T2WI and HT2WI than with single-source. However, 2 techniques showed similar lesion detectability.