Novel amide polar-embedded reversed-phase column for the fast liquid chromatography-tandem mass spectrometry method
P Herrero1, F Borrull, E Pocurull
1Department of Analytical Chemistry and Organic Chemistry, Universitat Rovira i Virgili, Sescelades Campus, Marcel-lí Domingo, s/n, 43007 Tarragona, Spain.
Abstract:
A fast chromatographic method has been developed that takes less than 5 min per run to determine five polyether ionophores with a novel amide polar-embedded reversed-phase column coupled to a triple quadrupole mass spectrometer. A comparison between Oasis HLB and Oasis MAX sorbents for the solid-phase extraction was done. Oasis HLB sorbent gave recoveries close to 90% and the repeatability (%RSD, 25-100 ng/L, n=3) of the method was less than 7% for all compounds in all matrices. The presence of polyether ionophores in environmental waters such as river water and sewage was investigated. Monensin and narasin were frequently determined in influent and effluent sewage at concentrations from 10 ng/L to 47 ng/L in influents and from 6 ng/L to 34 ng/L in effluents. In river waters, polyether ionophores were not detected in any sample.
More Related Videos
10:41Ion Exchange Chromatography (IEX) Coupled to Multi-angle Light Scattering (MALS) for Protein Separation and Characterization
Published on: April 5, 2019
05:46Identification of Pharmaceuticals in The Aquatic Environment Using HPLC-ESI-Q-TOF-MS and Elimination of Erythromycin Through Photo-Induced Degradation
Published on: August 1, 2018
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Analyte Adsorption and Distribution
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Types of Detectors