Published on: February 5, 2016
Stability of electrolyte determinations on the Siemens Advia 1800 analyzer
Rafael Garcia1, Geling Li, Zhenglong Wang
1Department of Pathology, SUNY Downstate Medical Center, Brooklyn, New York, NY, USA.
Background:
It is sometimes necessary for the laboratory to re-test samples for critical serum electrolyte levels. It is important to assure reproducibility of results when testing is performed on stored, refrigerated samples. We have tested the reproducibility of results for the critical electrolytes, Na, K, Cl and Ca, from ten randomly selected patients'sera over our maximum storage period of nine (9) days on the Siemens Advia 1800 analyzer. The ranges for each electrolyte were 131-150 meq/L (Na), 3.4-5.2 meq/L (K), 101-123 meq/L (Cl) and 7.3-9.9 mg/dL (Ca).
Methods:
We used ion-selective electrodes for Na, K and Cl and the ortho-cresolphthalein dye method for Ca.
Results:
We find that the reproducibility of determinations for all of these electrolytes was excellent, i.e. the coefficients of variation for each electrolyte determination for each patient were low.
Conclusion:
The methods of measurement for these electrolytes on the Advia 1800 are reliable and reproducible.
More Related Videos
Related Concept Videos
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Ion-Exchange Chromatography
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Electrolytes: van't Hoff Factor
EDTA: Indirect and Alkalimetric Titration
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...