Related Experiment Videos
Limitations of antenatal fetal heart rate monitors
G S Dawes1, M Moulden, C W Redman
1Nuffield Department of Obstetrics and Gynaecology, John Radcliffe Hospital, Oxford, England.
American Journal of Obstetrics and Gynecology
|January 1, 1990
Abstract:
Erroneous or doubtful decelerations in fetal heart traces were present in 111 of 1000 consecutive antenatal clinical records obtained by monitors with autocorrelation. The incidence was 20% in fetuses less than 30 weeks of gestational age. Their elimination reduced the number of "decelerative" records by 42%. Erroneous or doubtful accelerations were also present in 11% of records. These errors are caused by the fetal heart rate monitor and may contribute to the high intraobserver and interobserver variation on visual analysis. They can be detected by computer analysis.