Estimation performance of a reduced lead system during continuous 12-lead ECG ST-segment monitoring
Daniel Guldenring1, Dewar D Finlay, Stefan P Nelwan
1School of Computing and Mathematics and Computer Science Research Institute, University of Ulster, Belfast, UK. queldenring_daniel@gmx.de
Background:
Reduced lead systems utilizing patient-specific transformation weights have been reported to achieve superior estimates than those utilizing population-based transformation weights. We report upon the effects of ischemic-type electrocardiographic changes on the estimation performance of a reduced lead system when utilizing patient-specific transformation weights and population-based transformation weights.
Method:
A reduced lead system that used leads I, II, V2 and V5 to estimate leads V1, V3, V4, and V6 was investigated. Patient-specific transformation weights were developed on electrocardiograms containing no ischemic-type changes. Patient-specific and population-based transformations weights were assessed on 45 electrocardiograms with ischemic-type changes and 59 electrocardiograms without ischemic-type changes.
Results:
For patient-specific transformation weights the estimation performance measured as median root mean squared error values (no ischemic-type changes vs. ischemic-type changes) was found to be (V1, 27.5 μV vs. 95.8 μV, P<.001; V3, 33.9 µV vs. 65.2 µV, P<.001; V4, 24.8 μV vs. 62.0 μV, P<.001; V6, 11.7 μV vs. 51.5 μV, P<.001). The median magnitude of ST-amplitude difference 60 ms after the J-point between patient-specific estimated leads and actual recorded leads (no ischemic-type changes vs. ischemic-type changes) was found to be (V1, 18.9 μV vs. 61.4 μV, P<.001; V3, 14.3 μV vs. 61.1 μV, P<.001; V4, 9.7 μV vs. 61.3 μV, P<.001; V6, 5.9 μV vs. 46.0 μV, P<.001).
Conclusion:
The estimation performance of patient-specific transformations weights can deteriorate when ischemic-type changes develop. Performance assessment of patient-specific transformation weights should be performed using electrocardiographic data that represent the monitoring situation for which the reduced lead system is targeted.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring