Measured dose rate constant from oncology patients administered 18F for positron emission tomography
Brian Quinn1, Brian Holahan, Jean Aime
1Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY 10065, USA. quinnb@mskcc.org
Purpose:
Patient exposure rate measurements verify published patient dose rate data and characterize dose rates near 2-18-fluorodeoxyglucose ((18)F-FDG) patients. A specific dose rate constant based on patient exposure rate measurements is a convenient quantity that can be applied to the desired distance, injection activity, and time postinjection to obtain an accurate calculation of cumulative external radiation dose. This study reports exposure rates measured at various locations near positron emission tomography (PET) (18)F-FDG patients prior to PET scanning. These measurements are normalized for the amount of administered activity, measurement distance, and time postinjection and are compared with other published data.
Methods:
Exposure rates were measured using a calibrated ionization chamber at various body locations from 152 adult oncology patients postvoid after a mean uptake time of 76 min following injection with a mean activity of 490 MBq (18)F-FDG. Data were obtained at nine measurement locations for each patient: three near the head, four near the chest, and two near the feet.
Results:
On contact with, 30 cm superior to and 30 cm lateral to the head, the mean (75th percentile) dose rates per unit injected activity at 60 min postinjection were 0.482 (0.511), 0.135 (0.155), and 0.193 (0.223) μSv∕MBq h, respectively. On contact with, 30 cm anterior to, 30 cm lateral to and 1 m anterior to the chest, the mean (75th percentile) dose rates per unit injected activity at 60 min postinjection were 0.623 (0.709), 0.254 (0.283), 0.190 (0.218), and 0.067 (0.081) μSv∕MBq h respectively. 30 cm inferior and 30 cm lateral to the feet, the mean (75th percentile) dose rates per unit injected activity at 60 min postinjection were 0.024 (0.022) and 0.039 (0.044) μSv∕MBq h, respectively.
Conclusions:
The measurements for this study support the use of 0.092 μSv m(2)∕MBq h as a reasonable representation of the dose rate anterior from the chest of patients immediately following injection. This value can then be reliably scaled to the desired time and distance for planning and staff dose evaluation purposes. At distances closer than 1 m, a distance-specific dose rate constant of 0.367 μSv∕MBq h at 30 cm is recommended for accurate calculations. An accurate patient-specific dose rate constant that accounts for patient-specific variables (e.g., distribution and attenuation) will allow an accurate evaluation of the dose rate from a patient injected with an isotope rather than simply utilizing a physical constant.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
08:55A Practical Guide for the Production and PET/CT Imaging of 68Ga-DOTATATE for Neuroendocrine Tumors in Daily Clinical Practice
Published on: April 17, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET